Entry  Fri Sep 16 11:15:12 2011, Koji, Update, IOO, PZT1 situation P9151574.JPGP9161576.JPGP9161577.JPG
    Reply  Sat Sep 17 14:04:45 2011, Koji, Update, IOO, PZT1 driver in place P9171579.JPGP9171580.JPG
Message ID: 5447     Entry time: Sat Sep 17 14:04:45 2011     In reply to: 5431
Author: Koji 
Type: Update 
Category: IOO 
Subject: PZT1 driver in place 

The PZT driver is now in place. The actual PZTs are not connected yet!

It is accommodated on Ben's connector adapter board.

The panel has additional connectors now: two inputs and a power supply connector.

The supply voltage is +/-30V (actual maximum +/-40V), and the input range is +/-10V
which yields the output range of -5V to 30V. The gain of the amplifier is +2.

It is confirmed that the HV outputs react to the epics sliders although the PZT connector is not connected yet
so as not to disturb the locking activity.

When we engage the PZT connector, we should check the HV outputs with an oscilloscope to confirm they
have no oscillation with the capacitances of the PZTs together with the long cable.