Message ID: 5187     Entry time: Thu Aug 11 11:50:56 2011     In reply to: 5185     Reply to this: 5196
Author: Manuel , Ishwita 
Type: Update 
Category: PEM 
Subject: Calibration of Guralp and STS2 

 

 We just checked with a function generator the calibration of the ADC. We set a square wave with amplitude 1V. We measured the voltage with the oscilloscope and we found on the data viewer that one volt is 3208 counts. That's what we expected (+/- 10V for 16bits) but now we are more sure.

 