Wed Aug 10 14:27:39 2011, Ishwita , Manuel, Update, PEM, Calibration of Guralp and STS2
|
Wed Aug 10 14:35:39 2011, rana, Update, PEM, Calibration of Guralp and STS2
|
Wed Aug 10 15:17:39 2011, Ishwita, Manuel, Update, PEM, Calibration of Guralp and STS2
|
Thu Aug 11 09:39:25 2011, Ishwita, Manuel, Update, PEM, Calibration of Guralp and STS2
|
Thu Aug 11 11:50:56 2011, Manuel , Ishwita, Update, PEM, Calibration of Guralp and STS2
|
Thu Aug 11 16:15:59 2011, Ishwita, Manuel, Update, PEM, Calibration of Guralp and STS2
|
|
Message ID: 5187
Entry time: Thu Aug 11 11:50:56 2011
In reply to: 5185
Reply to this: 5196
|
Author: |
Manuel , Ishwita |
Type: |
Update |
Category: |
PEM |
Subject: |
Calibration of Guralp and STS2 |
|
|
We just checked with a function generator the calibration of the ADC. We set a square wave with amplitude 1V. We measured the voltage with the oscilloscope and we found on the data viewer that one volt is 3208 counts. That's what we expected (+/- 10V for 16bits) but now we are more sure.
|