Message ID: 5156     Entry time: Tue Aug 9 16:00:58 2011
Author: Jenny 
Type: Update 
Category: PSL 
Subject: Amplitude response of PZT 

AMresponsePZT.png

The top plot shows a sweep from 10 kHz to 5 MHz of the ratio of the voltage output of the PD detecting power from the NPRO laser beam and the RF source voltage (the magnitude of the complex transfer function). The black trace was taken with the laser beam blocked. For runs 2 and 3 I changed the laser temperature set point by 10 mK and 100 mK respectively to see if there was a significant change in the AM response. The bottom plots shows runs 2 and 3 compared to run 1 plotted in dB (to be explicit, i'm plotting 10 times the base 10 log of the magnitude of the ratio of two complex transfer functions). Changing the temperature seems to have only a minor effect on the output except at around 450kHz, where the response has a large peak in run 1 and much smaller peaks in runs 2 and 3. 

The traces in the top plot consist of 16 averages taken with a 300Hz IF bandwidth, 15 dBm source power (attenuated with a 6 dB attenuator) and with 20dB attenuation of the input power from the PD.

Next I'm going to probe a narrow band region where the response is low (2.0MHz or 2.4MHz perhaps) and choose a bandwidth for the dither frequency for the PDH locking.