Entry  Tue Aug 9 02:44:32 2011, Suresh, Update, SUS, ETMX free swinging data ETMX_freeswing.png
    Reply  Tue Aug 9 11:33:33 2011, kiwamu, Update, SUS, Re: ETMX free swinging data 
Message ID: 5153     Entry time: Tue Aug 9 11:33:33 2011     In reply to: 5150
Author: kiwamu 
Type: Update 
Category: SUS 
Subject: Re: ETMX free swinging data 

I believe that the 17 Hz broad structure on SIDE is just because of a bad rotational angle of the SIDE OSEM.

The same structure had been observed on the EMTY_UR, and the structure became narrower after we repositioned/rotated the OSEM yesterday.

My guess is that the SIDE OSEM is now in a place where the OSEM is quite sensitive to the bounce mode

and creating the broad structure due to a bi-linear coupling between the bounce mode and low frequency signals.

Quote from #5150

There is something defintely wrong with the side sensor.  It might be the electronics as it also has this problem with it slow channel readings (my previous elog today). 

ELOG V3.1.3-