40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 5063     Entry time: Fri Jul 29 18:43:02 2011
Author: Manuel, Ishwita 
Type: Update 
Category: PEM 
Subject: plugging seismometers to ADC 

[Manuel, Ishwita, Jenne, Jamie]

We changed the C1PEM model and the names of the C1:PEM channels.

We reinstalled the blue breakout box, since the purple one still didn't work.

So, now the AA board channels are connected as follows...

C1 = C1:PEM-SEIS_GUR1_X

C2 = C1:PEM-SEIS_GUR1_Y

C3 = C1:PEM-SEIS_GUR1_Z

C4 = C1:PEM-SEIS_GUR2_X

C5 = C1:PEM-SEIS_GUR2_Y

C6 = C1:PEM-SEIS_GUR2_Z

C7 = C1:PEM-SEIS_STS_1_X

C8 = C1:PEM-SEIS_STS_1_Y

C9 = C1:PEM-SEIS_STS_1_Z

C11 = C1:PEM-SEIS_STS_2_X

C12 = C1:PEM-SEIS_STS_2_Y

C13 = C1:PEM-SEIS_STS_2_Z

C14 = C1:PEM-SEIS_STS_3_X

C15 = C1:PEM-SEIS_STS_3_Y

C16 = C1:PEM-SEIS_STS_3_Z

C17 = C1:PEM-ACC_MC1_X

C18 = C1:PEM-ACC_MC1_Y

C19 = C1:PEM-ACC_MC1_Z

C20 = C1:PEM-ACC_MC2_X

C21 = C1:PEM-ACC_MC2_Y

C22 = C1:PEM-ACC_MC2_Z

Although the channels for all 3 STS-2 seismometers are made but only one is installed. So only Channels C1 to C9 are now in operation...

We checked the data from the plugged channels with the Dataviewer. We could see the peak whenever someone jumped in the lab. Even Kiwamu jumped and saw his signal.