The full characterization of POY11 is found in the PDF.
Resonance at 11.03MHz
Q of 7.6, transimpedance 1.98kOhm
shotnoise intercept current = 0.17mA (i.e. current noise of 7pA/rtHz)
Notch at 21.99MHz
Q of 56.2, transimpedance 35.51 Ohm
Notch at 55.20MHz
Q of 48.5, transimpedance 37.5 Ohm
|