Message ID: 4650     Entry time: Fri May 6 06:36:18 2011
Author: Suresh 
Type: Update 
Category: RF System 
Subject: PD DC signals at each port connected 

We now have the DC signal from three PDs available in the ADC channels 14,15 and 16.  The signals are from  REFL55, AS55 and POY photodiodes respectively.  As the DC signals on all the other PDs of the same port (REFL, AS and PO)  have the same information we do not need to monitor more than one DC PD at each port.

The LSC PD Interface Card, D990543 - Rev B, can take 4 PDs and provides the DC signals of the PDs on the connector P2 (the lower of the two) on the back plane of the chassis. An adaptor card, D010005-00, plugs into the back plane from the rear of the Eurorack and provides the four DC signals on two-pin lemo sockets.

I have connected the three DC signals from the relevant RF PDs (above) to a DC whitening filter, D990694-B-1 which is associated with the channels 9 to 16 of the ADC card.

The cables are in a bit of a mess right now as some of the PD power supply lines are too short to reach up the the Interface card in the top Eurocart. Steve and I plan to redo some of the cabling later today

 

 


 