Entry  Thu May 5 13:55:08 2011, kiwamu, Update, LSC, MI locking : calibration of BS and ITMs actuators MICH_OL.png
    Reply  Thu May 5 14:40:14 2011, Koji, Update, LSC, MI locking : calibration of BS and ITMs actuators 
       Reply  Thu May 5 15:28:23 2011, kiwamu, Update, LSC, Re: MI locking : calibration of BS and ITMs actuators 
Message ID: 4643     Entry time: Thu May 5 15:28:23 2011     In reply to: 4639
Author: kiwamu 
Type: Update 
Category: LSC 
Subject: Re: MI locking : calibration of BS and ITMs actuators 

They are the DC responses.

I put the resonant frequencies that Leo reported in the wiki to obtain the DC response.

The resonant frequencies I used are :

  f_BS = 0.957 Hz

  f_ITMX = 0.966 Hz

  f_ITMY = 0.988 Hz

Also I assumed that all the Q-values are 5 due to the damping.

Quote:

I've got confused

1) Are these the DC responses of the coils? If that is true, we need to specify the resonant frequency of each suspension to get the AC response.

2) Are these the AC responses well above the resonant freqs? In that case, The responses should be x.xxx / f^2 [m/counts]

Quote:

 BS = 3.69e-08 [m/counts]
 ITMX = 8.89e-09 [m/counts]
 ITMY = 9.22e-09 [m/counts]

 