40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Mar 30 19:43:33 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching 
    Reply  Wed Mar 30 19:54:19 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching laser_output_non_circular.png
       Reply  Wed Mar 30 20:08:34 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching Beam_Matching_02c_Vertical.pngBeam_Matching_02c_Horizontal.png
          Reply  Wed Mar 30 20:14:17 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching 
             Reply  Wed Mar 30 20:31:30 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching 
                Reply  Wed Mar 30 20:50:43 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching After_Faraday.pngAfter_Faraday_and_Rotated_Lens.png
                   Reply  Wed Mar 30 21:21:15 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching Oven_Lens_Solution_66mm.pngMatching_Into_Green_Oven_zoomed_out.pngMatching_Into_Green_Oven_zoomed_in.png
                      Reply  Thu Mar 31 02:59:49 2011, Koji, Configuration, Green Locking, The wonderful world of mode-matching 
                         Reply  Thu Mar 31 14:10:00 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching P3310028.JPGP3310029.JPG
                         Reply  Thu Mar 31 15:23:14 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching Zoomed_out_fitting01.pngZoomed_in_fitting01.png
                            Reply  Mon Apr 4 14:20:32 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching Waist_Fits_from_laser.pngWaist_Fits_Bench_Position.png
Message ID: 4467     Entry time: Wed Mar 30 20:14:17 2011     In reply to: 4466     Reply to this: 4468
Author: Bryan 
Type: Configuration 
Category: Green Locking 
Subject: The wonderful world of mode-matching 

Additional:

I fired up some old waistplotter routines, and set the input conditions as the measured waist after the lens and used that to work out what the input waist is at the laser. It may not be entirely accurate, but it /will/ be self consistent later on.

 

Vertical waist      = 105.00 um at 6.282 cm after laser output (approx)

Horizontal waist = 144.63 um at 5.842 cm after laser output (approx)

 

Definitely astigmatic.

 