40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Mar 30 19:43:33 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching 
    Reply  Wed Mar 30 19:54:19 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching laser_output_non_circular.png
       Reply  Wed Mar 30 20:08:34 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching Beam_Matching_02c_Vertical.pngBeam_Matching_02c_Horizontal.png
          Reply  Wed Mar 30 20:14:17 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching 
             Reply  Wed Mar 30 20:31:30 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching 
                Reply  Wed Mar 30 20:50:43 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching After_Faraday.pngAfter_Faraday_and_Rotated_Lens.png
                   Reply  Wed Mar 30 21:21:15 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching Oven_Lens_Solution_66mm.pngMatching_Into_Green_Oven_zoomed_out.pngMatching_Into_Green_Oven_zoomed_in.png
                      Reply  Thu Mar 31 02:59:49 2011, Koji, Configuration, Green Locking, The wonderful world of mode-matching 
                         Reply  Thu Mar 31 14:10:00 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching P3310028.JPGP3310029.JPG
                         Reply  Thu Mar 31 15:23:14 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching Zoomed_out_fitting01.pngZoomed_in_fitting01.png
                            Reply  Mon Apr 4 14:20:32 2011, Bryan, Configuration, Green Locking, The wonderful world of mode-matching Waist_Fits_from_laser.pngWaist_Fits_Bench_Position.png
Message ID: 4466     Entry time: Wed Mar 30 20:08:34 2011     In reply to: 4465     Reply to this: 4467
Author: Bryan 
Type: Configuration 
Category: Green Locking 
Subject: The wonderful world of mode-matching 

Step 2: Getting the beam through the Faraday isolator (FI).

Started out with an f=100mm lens at position 32,T on the bench which gave a decent looking waist of order 100 um in the right sort of position for the FI, but after checking the FI specs, it's limited to 500W/cm^2. In other words, if we have full power from the laser passing into it we'd need a beam width of more than 211 um. Solution? Use an f=150mm lens instead and don't put the FI at the waist. I normally don't put a FI at a waist anyway, for assorted reasons - scattering, thermal lensing, non-linear magnetic fields, the sharp changing of the field components in an area where you want as constant a beam as possible.  Checked with others to make sure they don't do things differently around these parts… Koji says it doesn't matter as long as it passes cleanly through the aperture. So… next step is inserting the Faraday.

The beam profiles in vertical and horizontal around the FI position with the f=150mm lens in place are attached. Note that the FI will be going in at around 0.56m.

Beam_Matching_02c_Vertical.pngBeam_Matching_02c_Horizontal.png

 

 

 