Message ID: 4424     Entry time: Tue Mar 22 16:39:51 2011
Author: kiwamu 
Type: Update 
Category: Green Locking 
Subject: comaprator installed : 80 pm residual displacement 

 A comparator has been installed before the MFDs (mixer-based frequency discriminator) to eliminate the effect from the amplitude fluctuation (i.e. intensity noise).

As a result we reached an rms displacement of 580 Hz or 80 pm.

 

(differential noise measurement)

noise_curve20110322.png

  Here is the resultant plot of the usual differential noise measurement.

The measurement has been done when the both green and red lasers were locked to the X arm.

In the blue curve I used only MFD. In the black curve I used the combination of the comparator and the MFD.

Noise below 3 Hz become lower by a factor of about 4, resulting in a better rms integrated from 40 Hz.

Note that the blue and the black curve were taken while I kept the same lock.

A calibration was done by injecting a peak at 311 Hz with an amplitude of 200 cnt on the ETMX_SUS_POS path.

 

(installation)

  Yesterday Koji modified his comparator circuit such that we can take a signal after it goes thorough the comparator.

The function of this comparator is to convert a sinusoidal signal to a square wave signal so that the amplitude fluctuation doesn't affect the frequency detection in the MFD.

I installed it and put the beat-note signal to it. Then the output signal from the comparator box is connected to the MFDs.

The input power for the comparator circuit has been reduced to -5 dBm so that it doesn't exceeds the maximum power rate.