40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Mar 8 06:29:06 2011, kiwamu, Update, Green Locking, Intensity noise setup 
    Reply  Wed Mar 9 04:46:13 2011, kiwamu, Update, Green Locking, more intensity noise measurement intensity_setup.pngIntensityNoise.png
       Reply  Wed Mar 9 18:17:11 2011, kiwamu, Update, Green Locking, Intensity noise coupling diff_noise.pngIN_TF.png
          Reply  Thu Mar 10 14:06:54 2011, kiwamu, Update, Green Locking, Intensity noise limits the beatnote sensitivity diff_noise_Mar8.pngIN_TF_complete.png
             Reply  Thu Mar 10 14:22:58 2011, kiwamu, Update, Green Locking, Intensity noise limits the beatnote sensitivity 
                Reply  Thu Mar 10 14:29:05 2011, Koji, Update, Green Locking, Intensity noise limits the beatnote sensitivity 
                Reply  Thu Mar 10 14:30:53 2011, rana, Update, Green Locking, Intensity noise limits the beatnote sensitivity 
                   Reply  Thu Mar 10 16:00:53 2011, Aidan, Joe, Update, Green Locking, Intensity stabilization loop using beatnote DC. P1000313.jpgP1000314.jpgP1000315.jpgGREEN_ISS_LOOP.pdfScreenshot-C1ALS_OVERVIEW.adl.png
                      Reply  Thu Mar 10 21:45:34 2011, rana, Update, Green Locking, Intensity stabilization loop using beatnote DC. 
Message ID: 4398     Entry time: Thu Mar 10 14:22:58 2011     In reply to: 4397     Reply to this: 4399   4400
Author: kiwamu 
Type: Update 
Category: Green Locking 
Subject: Intensity noise limits the beatnote sensitivity 

The next steps we should do are :

    - to investigate a cause of the intensity fluctuation
          * end green laser
          * suspensions' angular motions
          * doublecheck the RIN contribution if it's from the PSL or the X arm in the beatnote RFPD to make sure the RIN is dominated by the X arm transmitted light
  
    - to think about how to make the system insensitive to the intensity noise
          - bring the beat frequency to the zero cross point of the MFDs ?
          - PLL ?

Quote:

We are limited by the intensity noise of the X arm transmitted green light.