Today I was working on RF distribution box.
So far I almost finished to electronically isolate voltage regulators from the box wall by inserting mica sheet, sleeve, and washers.
The problem I found is the resistance between wall and the voltage regulator is order of M ohms
I checked my isolation (mica sheet and sleeve and washer) but there is no problem there.
But I found that the power switch is not completely isolated from the wall.( around 800 kohm)
and that the resistance between the regulator and the wall is smaller for the regulator closer to the power switch
and greater for the regulator less closer to it.
So I think we need to put washer or sleeve to isolate the powersitch electronically from the box wall.
Suresh or I will fix this problem