Today's CDS problems |
|
|
This is just a listing of CDS problems I still notice today:
-
MC2-MCL button was left ON due to BURT failure. This, of course, screws up our Green locking investigations because of the unintended feedback. Please fix the BURT/button issue.
- The GCV - FB0 status is RED. I guess this means there's something wrong? Its really a bad idea to have a bunch of whited out or falsely red indicators. No one will ever use these or trust these in the future.
- MC1/2/3 Lockins are all white. Also, the MODE switches for the dewhitening are all white.
- Is the MC SIDE coil dewhitening filter synced with anything? It doesn't seem to switch anything. Maybe the dewhite indicators at the top right of the SUS screens can be made to show the state of the binary output instead of just the digital filter
- MC WFS is all still broken. We need a volunteer to take this on - align beams, replace diodes, fix code/screens.
|