Message ID: 4143     Entry time: Wed Jan 12 17:22:47 2011
Author: Suresh 
Type: Configuration 
Category: Locking 
Subject: MC demod phase adjusted to minimise the I output 

[Koji, Suresh]

We wanted to check and make sure that the relative phase of the two inputs ( local oscillator and photodiode signal ) to the demod board is such that the Q output is maximised.   We displayed the I and  Q signals on the oscilloscope in XYmode with I along the X direction.  If Q is maximised (and therefore I is minimised) the oscillocope trace would be perfectly vertical since all the signal would be in Q and none in I. Initially we noted that the trace was slightly rotated to the CCW of the vertical and that a short cable was present in the PD input line.  Removing this rotated the trace CW and made it pretty much vertical.  The screen shot of the oscilloscope is below.