This is looking better, but the fit data for the TF should be plotted along with the data. The data should be made up of points and the fit a line.
For the fit, we should have the Q of the main resonance as well as the peak height of the main resonance and the values of the gain at the notch frequencies.
Also the peak as well as the notches should have the frequencies fit for and labeled. In principle, you can make the plot on the wiki have all of the data. Then in the end we can print the plot in a small size and glue it to the PD's backside. |