Problem:
We couldn't set testpoints on the c1ioo machine.
Cause:
Awgtpman was getting into some strange race condition. Alex added an additional sleep statement and shifted boot order slightly in the rc.local file. This apparently is only a problem on c1ioo, which is a Sun X4600. It was using up 100% of a single CPU for the awgtpman process.
Status:
We now have c1ioo test points working.
Future:
Need to examine the startc1ioo script and see if needs a similar modification, as that was tried at several points but yielded a similar state of non-functioning test points. For the moment, reboot of c1ioo is probably the best choice after making modifications to the c1ioo or c1x03 models.
Current CDS status:
MC damp |
dataviewer |
diaggui |
AWG |
c1ioo |
c1sus |
c1iscex |
RFM |
Sim.Plant |
Frame builder |
TDS |
|
|
|
|
|
|
|
|
|
|
|
|