Message ID: 3830
Entry time: Sat Oct 30 14:35:43 2010
In reply to: 3829
Reply to this: 3838
|
Author: |
Koji |
Type: |
Summary |
Category: |
CDS |
Subject: |
CDS time delay measurement |
|
|
Unsatisfactory.
Neglecting the digital anti-imaging filter makes the discrepancy. You must take into account your digital filter twice.
I attached the slides I made during my visit for March LVC '09. P.5 would be useful.
Quote: |
Result:
[time delay of the CDS] (left, middle)
The time delay gets larger with frequency. The time delay seems to be -175 usec at DC.
However, the gain seems a little different from my expectation(feCoeff4x). So, there are maybe other filters I don't know.
I neglected TF of upsampling this time.
|
|
|
|