Entry  Fri Oct 22 19:53:27 2010, yuta, Summary, CDS, conversion failure in digital filters 
    Reply  Sat Oct 23 03:36:05 2010, yuta, Update, CDS, fixed filters for C1SUS, C1RMS, C1MCS 
Message ID: 3769     Entry time: Sat Oct 23 03:36:05 2010     In reply to: 3765
Author: yuta 
Type: Update 
Category: CDS 
Subject: fixed filters for C1SUS, C1RMS, C1MCS 

(Joe, Yuta)

Summary:

 This Monday, MC suspension damping got something wrong.
 We started to check filters and found that digital filters were wrong because of mis-conversion from old filter files to new files.
 We converted the file again, and with mutual understanding between Foton and us, we finally got correct filters(I hope!).

What we did:
 1. Merged filter files in old /cvs/cds/caltech/chans/ directory into C1SUS.txt(BS,ITMX,ITMY), C1RMS.txt(PRM,SRM), C1MCS.txt(MC123).

 2. Rebuilt the RT models in order to get a correct filter file header(they have list of filter modules).

 3. Concatenate the header with filter design part which we got from step1.

 4. Replaced 'N 2048' with 'N 16384'
   It replaces sampling rate of "XXSEN"s.

Basically, step1-4 was the same with what we did last time. We didn't changed the fitler coefficients, so Foton somehow changed the original filter design.
So, this time, we

 5. Deleted coefficients like we did on Tuesday (see elog #3774).

But Foton couldn't read the file correctly this time. Foton seemed to be unbeatable.
Even if we replaced the sampling rate, Foton kept saying 2048! (This is maybe because Foton's default value is 2048Hz. Everytime Foton notice some editting in the file, he destroys everything. He hates editting)
The problems were always associated with the sampling rate, so

 6. Got back to step4 and undo-ed the replacement.

 7. Foton could read it this time, so I changed the sampling rate one by one using Foton GUI.

 8. Checked filters using Foton's Bode Plot.(Not for all, but some that had problem before)

 9. Splitted SDSEN filters to SDSEN, SUSSIDE, and SDCOIL.

 10. Put some missing whitening filters, and 28HzELPs.
   BS, PRM, SRM didn't have any 28HzELP for SDCOIL.
   ITMX and ITMY SDCOIL had SimDW/InvDW which doesn't make sense(SIDEs don't have analog DW). So, I deleted and replaced with 28HzELP.

F2A issue:
 We failed in sending F2A filters to new filter files.
 These are a little bit complicated because TO_COIL_X_X filters were named ULPOS,URPOS etc before.
 Also, MC3 didn't have any F2As, so maybe we should but the same F2As as MC1/2.
 Note that every F2As are different, and TO_COIL matrix have UL,UR,LL,LR order(not same as INMATRIX).
 Also, SRM had f2pv instead of F2A!

Next work:
 - Check whole filters by actually measuring transfer function between SENs and COILs.
 - Damp MC suspentions, and lock MC.
 - Measure openloop TF and compare with the designed.


How do you read a Foton filter file:
 When you open up a Foton filter file, you see filters like this.

################################################################################
### modulename                                                               ###
################################################################################
# SAMPLING modulename samplingrate
# DESIGN   modulename n filterdesign
# DESIGN   modulename n filterdesign
###                                                                          ###
modulename   n xy z      v      w filtername                        gain    a1     a2    b1     b2
modulename   n xy z      v      w filtername                        gain    a1     a2    b1     b2
                                                                            a1     a2    b1     b2
n: filter number
 0 for FM1, 1 for FM2, ... , 9 for FM10

x: Input Switching setting
 1 Always On
 2 Zero History

y: Output Switching setting
 1 Immediately
 2 Ramp
 3 Input Crossing
 4 Zero Crossing

z: number of filters cascaded.

v: if y=2, (Ramp Time(sec))*(samplingrate)
   if y=3 or 4, Tolerance

w: (Timeout(sec))*(samplingrate)

 Note that v and w are changed when sampling rate is changed.

 Transfer function will be;
  H(1/z)=G*(1+b1/z+b2/z/z)/(1+a1/z+a2/z/z)
  z=exp(s/fs)

 where fs is the sampling frequency.

Reference:
 Kiwamu Izumi: "Notes about Digital Filters," http://tamago.mtk.nao.ac.jp/izumi/green/DigitalFilter.pdf