Entry  Fri Oct 22 19:53:27 2010, yuta, Summary, CDS, conversion failure in digital filters 
    Reply  Sat Oct 23 03:36:05 2010, yuta, Update, CDS, fixed filters for C1SUS, C1RMS, C1MCS 
Message ID: 3765     Entry time: Fri Oct 22 19:53:27 2010     Reply to this: 3769
Author: yuta 
Type: Summary 
Category: CDS 
Subject: conversion failure in digital filters 

(Rana, Joe, Yuta)

We now understand that we never succeeded in converting old fiter files to new filter files.

For example, we just changed the sampling rate with coefficients remained and foton confused, or we forgot to rename some of the module names(ULSEN -> SUS_MC1_ULSEN) ......

This is why we sometimes damped and sometimes didn't, depending on filter switches. New filter files has been always wrong.

So, we started to convert them again.
We have to figure out how to convert the files that Foton accepts correctly.