40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Thu Oct 7 14:39:50 2010, yuta, Update, PSL, measured PMC's laser power-output relation
|
Thu Oct 7 15:05:46 2010, Koji, Update, PSL, measured PMC's laser power-output relation
|
Thu Oct 7 16:34:06 2010, yuta, Update, PSL, measured PMC's laser power-output relation
|
Thu Oct 14 12:53:29 2010, yuta, Update, PSL, mesured PMC's visibility vs power relation  
|
|
Message ID: 3667
Entry time: Thu Oct 7 14:39:50 2010
Reply to this: 3669
|
Author: |
yuta |
Type: |
Update |
Category: |
PSL |
Subject: |
measured PMC's laser power-output relation |
|
|
(Rana, Yuta)
Motivation:
We wanted to see thermal effects on the PMC.
What I did yesterday:
Changed the current of the NPRO from 2A to 0.8A and measured the power of the reflected/transmitted light from the PMC when locked.
I also measured the power of the reflected light when PMC is not locked (It supposed to be proportional to the output power of the laser).
Result:
Attached. Hmmmm......
At several points of the laser current, I could'nt lock the PMC very well. The power of the reflected/transmitted light depend on the offset voltage of the PZT.
When the laser power was weak(~<0.9A), the power of reflected/transmitted light changed periodically(~ several minutes). |
|
|