40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Sep 27 00:33:50 2010, rana, Update, PSL, High Voltage Driver added to TTFSS -> NPRO 
    Reply  Fri Oct 1 21:34:14 2010, rana, tara, Update, PSL, High Voltage Driver added to TTFSS -> NPRO NPROTF.png2010_10_01.png
       Reply  Mon Oct 4 06:47:46 2010, rana, tara, Update, PSL, High Voltage Driver added to TTFSS -> NPRO 
Message ID: 3640     Entry time: Fri Oct 1 21:34:14 2010     In reply to: 3610     Reply to this: 3641
Author: rana, tara 
Type: Update 
Category: PSL 
Subject: High Voltage Driver added to TTFSS -> NPRO 

Quote:

We added the Thorlabs HV Driver in between the FSS and the NPRO today. The FSS is locking with it, but we haven't taken any loop gain measurements.

This box takes 0-10 V and puts out 0-150 V. I set up the FSS SLOW loop so that it now servos the output of FAST ot be at +5V instead of 0V. This is an OK

temporary solution. In the future, we should add an offset into the output of the FSS board so that the natural output is 0-10 V.

I am suspicious that the Thorlabs box has not got enough zip to give us a nice crossover and so we should make sure to measure its frequency response with a capacitive load.

 

 We measured the Thorlabs HV Driver's TF today. It is quite flat from 1k to 10k before going up to 25 dB at 100k,

and the response does not change with the DC offset input.

 

The driver is used for driving the NPRO's PZT which requires higher voltage than that of the previous setup.

We need to understand how the driver might effect the FSS loop TF, and we want to make sure that the driver

will have the same response with DC input offset.

 

Setup

 

We used SR785 to measure the TF. Source ch was split by a T, one connected to Driver's input, another one connected to the reference (ch A). See fig2.

The driver output was split by another T. One output was connected to NPRO,

another was connected to a 1nF capacitor in a Pomona box, as a high pass filer (for high voltage), then to the response (ch B)

 The source input is  DC offset by 2V which corresponds to 38 V DC offset on the driver's output.

The capacitance of the PZT on the NPRO is 2.36 nF, as measured by LC meter.

 

 The result shows that the driver's TF is flat from 1k to 10k, and goes up at higher frequency, see fig1.

 

The next step is trying to roll of the gain at high frequency for PZT. A capacitor connected to ground might be used to roll off the frequency of the driver's output.

We will inspect the TF at higher frequency (above 100 kHz) as well.

            

Attachment 1: NPROTF.png  23 kB  Uploaded Fri Oct 1 23:03:27 2010  | Hide | Hide all | Show all
NPROTF.png
Attachment 2: 2010_10_01.png  15 kB  Uploaded Fri Oct 1 23:58:39 2010  | Show | Hide all | Show all
ELOG V3.1.3-