Message ID: 363     Entry time: Fri Mar 7 00:47:54 2008
Author: rana 
Type: Configuration 
Category: PEM 
Subject: Ranger SS-1 
Yesterday evening around 7:30 PM, I changed the Ranger seismometer from a
vertical to a horizontal seismometer. To do this I followed the instructions
in the manual.
1) Lock it down.
2) Turn it sideways. Use the leveling screws to center the bubble level.
3) Carefully loosen the hanger rod and release slowly the tension to allow
   the mass to recenter.
4) Look through the little viewhole next to the rod to make the white lines
   line up. This means the mass is centered.
5) Look at the output on a scope. It should be freely moving with a ~1 sec.
   period.

The attached plot shows the before and after spectra.
Attachment 1: ss1.pdf  53 kB  | Hide | Hide all
ss1.pdf