Entry  Thu Sep 30 17:11:01 2010, alex i, Update, CDS, DAQ system update 
    Reply  Thu Sep 30 21:55:31 2010, rana, Update, CDS, DAQ sys update Untitled.png
Message ID: 3629     Entry time: Thu Sep 30 17:11:01 2010     Reply to this: 3631
Author: alex i 
Type: Update 
Category: CDS 
Subject: DAQ system update 

The frame builder is timed from the Symmetricom GPS card now, which is getting the IRIGB timecode from the freq. distribution amplifier (from the VME GPS receiver card).

I have adjusted the GPS seconds to match the real GPS time and the DTT seems to be happy: sweeping MC2 MCL filter module produces nice plot.

Test points are working on SUS.

Excitations are working on SUS.

I am leaving the frame builder running and acquiring the data.

 

Alex

 