Entry  Fri Sep 24 23:47:10 2010, koji, tara, Update, Electronics, testing TTFSS 
    Reply  Sat Sep 25 19:01:13 2010, Koji, Update, Electronics, testing TTFSS 
Message ID: 3607     Entry time: Fri Sep 24 23:47:10 2010     Reply to this: 3608
Author: koji, tara 
Type: Update 
Category: Electronics 
Subject: testing TTFSS 

  Q3, a PZT2222A transistor, on D0901846 is replaced by a GE-82. However, the board is still not fully function.

 

Since Q3, PZT2222A, was broken, I went to Wilson house and got some SP3904's for replacement. But somehow, I broke it during

installation, and did not notice it, and resumed the test. When I got to test 8 on the list, the TTFSS did not work as specified.

Koji checked and found out that -15V, Nref, Vref voltages output did not work correctly. So the SP3904 I installed was removed

and replaced with another SP3904 by Koji, and Vref is working. 

Q4 transistor is broken as well and it was replaced by GE 82.

Q1 might be broken too since -15V out is not working.

I'll go to Wilson house to get more transistors next week.

 

After the broken parts have been replaced, I have to make sure that I separate the power supply board from the rest of the circuit and

check if all V outputs are  working, then reconnect the board and check if the current input is reasonable before resume the test.

I hope the wrong input voltage problem today wouldn't damage anything else.

 