Subject: PMC line width 

The attached plots show the PMC cavity line width measurement with 1 mW and 160 mW into the PMC. The two curves on each plot are the PMC transmitted power and the ramp of the fast input of the NPRO. The two measurements are consistent within errors - a few %. The PMC line width  3.5 ms (FWHM) x 4 V / 20 ms (slope of the ramp) x 1.1 MHz / V (NPRO fast actuator calibration from Innolight spec sheet) = 0.77 MHz.

Here is the output of the calculation using Malik Rakhmanov code:

 

modematching =  8.4121e-01

transmission1 =   2.4341e-03

transmission2 =   2.4341e-03

transmission3 =   5.1280e-05

averageLosses =  6.1963e-04

visibility =  7.7439e-01

Here are the inputs for the calculation in the param.m:

 

fw = 0.77e6;                % width of resonance (FWHM) in Hz

Plas = 0.164;                % power into the PMC in W

 

% the following number refer to the in-lock cavity state

 

Pref = 0.037;                % reflected power in W

Ptr = 0.0712;                 % transmitted power in W

Pleak = 0.0015;              % power leaking from back of PMC in W

 

 