40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Feb 19 10:14:13 2008, steve, Update, VAC, rga logging needs help 
    Reply  Thu Feb 28 13:04:59 2008, rob, Update, VAC, rga logging working again 
       Reply  Mon Mar 3 09:25:33 2008, steve, Update, VAC, rga scan logging is working now rgalogworks.jpgcc1.jpg
Message ID: 351     Entry time: Mon Mar 3 09:25:33 2008     In reply to: 344
Author: steve 
Type: Update 
Category: VAC 
Subject: rga scan logging is working now 

Quote:

Quote:
The rga head and controller are running fine, but the data logging is not.


It should run tonight at 1:25 AM. To get the cron job to work properly on op340m, I had to make wrapper sh script which defines the perl library before calling the actual RGAlogger script.




Thanks to Rob, it is working !

The baked, calibrated rga head
model# M206M, s/n #c128035 was reinstalled at the 40m ifo on Feb. 8, 2008
Faraday mode dwell time was increased from 2 to 16 sec
Rga head temp at top silver gaskit 45.2 C
The noise floor is at 1E-12 Torr

There is more detail in logbook VMI-14 p 107

pd65-m-d119
Attachment 1: rgalogworks.jpg  82 kB  | Show | Show all
Attachment 2: cc1.jpg  107 kB  Uploaded Mon Mar 3 10:55:47 2008  | Show | Show all
ELOG V3.1.3-