Result of the accelerometer measurement
Introduction
We wanted to characterize the PSL table before the work before its lifting up.
We put a set of three-axis Wilcoxon accelerometers on the ground and another set on the PSL table through the weekend.
Result
- The data at 9th Aug 00:00(UTC) is used. This was Sunday 5PM in the local time.
- The freq resolution was 0.01Hz. The # of avg was 50.
- The accelerometer signals were calibrated by the value 1.2e-7 V/(m/s^2). We use this absolute value of the spectrum for the comparison purpose.
- The accelerometers were aligned to North(X), East(Y), and Up(Z). There was the coherence observed from 2~20Hz.
The transfer functions are valid only this frequency region although we still can set the lower bound of them.
- The transfer functions in the horizontal directions show huge peaks at around 20Hz. The Q of the peaks are ~30 to ~100.
The vertical transfer function shows somewhat lower peak at around 50Hz with Q of ~10.
Some thoughts
- The low resonant freq and the high Q of the horizontal mode comes from the heaviness of the table.
- We are going to raise the table. This will usually mean that we get the lower resonant freq. This is not nice.
- So, the decision to use 6 tripods rather than 4 was right.
- The steel tripods are expected to give both more rigidity and more damping than the chep-looking hollow Newport legs.
- Concrete grouting of the tripods will also lower the effective height and will benefit for us.
|