Message ID: 3243     Entry time: Mon Jul 19 13:51:09 2010
Author: kiwamu 
Type: Update 
Category: CDS 
Subject: TIming card at X end 

[Joe, Kiwamu]

 

 The timing slave in the IO chassis on the new X end was not working with symptoms of no front "OK" green light, no "PPS" light, 3.3V testpoint not working and  ERROR testpoint bouncing between 5-6V.

We took out the timing slave from the X end IO chassis put in to the new Y end IO chassis .

It worked perfectly there. We took the working one from Y end put in the X end IO chassis.

We slowly added cables. First we added power , it worked fine and we saw green "OK" light. Then we added 1PPS signal by a fiber and it also worked.

We turned everything off and then we added 40pin IPC cable from the chassis and infiniband cable from the  computer.

When we turned ON it we didn't see the green light.

This means something in the computer configuration might be wrong not in the timing card, we now are trying to make contact with Alex.

We are comparing the setup of the C1SCX  machine and the working C1ISCEX machine.