Message ID: 3031     Entry time: Wed Jun 2 03:50:03 2010     In reply to: 3029     Reply to this: 3043
Author: Koji 
Type: Update 
Category: IOO 
Subject: mode measurement of new input optics 

Just note that MMT1 has RoC of -5m (negative!). This means that it is a lens with f=-2.5 m,

 