Entry  Sun May 16 12:51:08 2010, kiwamu, Update, Green Locking, reflected beam at PD 
    Reply  Sun May 16 19:25:45 2010, Koji, Update, Green Locking, reflected beam at PD 
       Reply  Thu May 20 13:29:40 2010, kiwamu, Update, Green Locking, mode profile at PPKTP crystal PPKTPmode.png
          Reply  Thu May 20 14:18:59 2010, kiwamu, Update, Green Locking, mode profile of 40m cavity MMT40mcavity.png
             Reply  Wed May 26 04:14:21 2010, kiwamu, Update, Green Locking, locked OLTF_endPDH.png
                Reply  Wed May 26 14:38:02 2010, Koji, Update, Green Locking, locked 
                   Reply  Thu May 27 02:22:24 2010, kiwamu, Update, Green Locking, more details stimes.pngspectrum.pngETMX_small.png
                      Reply  Thu May 27 09:43:50 2010, rana, Update, Green Locking, more details 
                         Reply  Mon Jul 12 13:16:53 2010, kiwamu, Update, Green Locking, PZT feedback at X end arm_day_night.png
                      Reply  Wed Jun 9 11:13:11 2010, kiwamu, Update, Green Locking, lock with PDH box 
                         Reply  Fri Jun 11 11:54:42 2010, kiwamu, Update, Green Locking, end PDH with thermal feedback LPF.pngthermal_feedback.png
                            Reply  Fri Jun 11 13:32:28 2010, Koji, Update, Green Locking, end PDH with thermal feedback 
                         Reply  Sun Jun 20 18:28:25 2010, kiwamu, Update, Green Locking, Re: lock with PDH box 
                Reply  Wed Jul 21 19:11:18 2010, Dmass, Update, Green Locking, locked 
Message ID: 2997     Entry time: Thu May 27 02:22:24 2010     In reply to: 2992     Reply to this: 2999   3059
Author: kiwamu 
Type: Update 
Category: Green Locking 
Subject: more details 

 Here are some more plots and pictures about the end PDH locking with the green beam. 

-- DC reflection

 I expected that the fluctuation of the DC reflection had 1% from the resonant state to the anti-resonant state due to its very low finesse.

This values are calculated from the reflectivity of ETM measured by Mott before (see the wiki).

In my measurement I obtained  DC reflection of V_max=1.42 , V_min=1.30  at just after the PD.

These numbers correspond to 7.1% fluctuation. It's bigger than the expectation.

I am not sure about the reason, but it might happen by the angular motion of test masses (?)

 

--- time series

Here is a time series plot. It starts from openloop state (i.e. feedback disconnected).

At t=0 sec I connected a cable which goes to the laser pzt, so now the loop is closed.

You can see the DC reflection slightly decreased and stayed lower after the connection.

The bottom plot represents the feedback signal measured before a sum amp. which directly drives the pzt.

stimes.png

 

 

-- length fluctuation  

One of the important quantities in the green locking scheme is the length fluctuation of the cavity.

It gives us how much the frequency of the green beam can be stabilized by the cavity. And finally it will determine the difficulty of PLL with the PSL.

I measured a spectrum of the pzt driving voltage [V/Hz1/2] and then converted it to a frequency spectrum [Hz/Hz1/2].

I used the actuation efficiency of 1MHz/V for the calibration, this number is based on the past measurement.

spectrum.png

RMS which is integrated down to 1Hz  is 1.6MHz.

This number is almost what I expected assuming the cavity swings with displacement of x ~< 1um.

 

-- flashing

A picture below is a ETMx CCD monitor.

One of the spot red circled in the picture blinks when it's unlocked. And once we get the lock the spot stays bright.

ETMX_small.png

 