Entry  Sun May 16 12:51:08 2010, kiwamu, Update, Green Locking, reflected beam at PD 
    Reply  Sun May 16 19:25:45 2010, Koji, Update, Green Locking, reflected beam at PD 
       Reply  Thu May 20 13:29:40 2010, kiwamu, Update, Green Locking, mode profile at PPKTP crystal PPKTPmode.png
          Reply  Thu May 20 14:18:59 2010, kiwamu, Update, Green Locking, mode profile of 40m cavity MMT40mcavity.png
             Reply  Wed May 26 04:14:21 2010, kiwamu, Update, Green Locking, locked OLTF_endPDH.png
                Reply  Wed May 26 14:38:02 2010, Koji, Update, Green Locking, locked 
                   Reply  Thu May 27 02:22:24 2010, kiwamu, Update, Green Locking, more details stimes.pngspectrum.pngETMX_small.png
                      Reply  Thu May 27 09:43:50 2010, rana, Update, Green Locking, more details 
                         Reply  Mon Jul 12 13:16:53 2010, kiwamu, Update, Green Locking, PZT feedback at X end arm_day_night.png
                      Reply  Wed Jun 9 11:13:11 2010, kiwamu, Update, Green Locking, lock with PDH box 
                         Reply  Fri Jun 11 11:54:42 2010, kiwamu, Update, Green Locking, end PDH with thermal feedback LPF.pngthermal_feedback.png
                            Reply  Fri Jun 11 13:32:28 2010, Koji, Update, Green Locking, end PDH with thermal feedback 
                         Reply  Sun Jun 20 18:28:25 2010, kiwamu, Update, Green Locking, Re: lock with PDH box 
                Reply  Wed Jul 21 19:11:18 2010, Dmass, Update, Green Locking, locked 
Message ID: 2960     Entry time: Thu May 20 14:18:59 2010     In reply to: 2959     Reply to this: 2988
Author: kiwamu 
Type: Update 
Category: Green Locking 
Subject: mode profile of 40m cavity 

The mode profile of the green beam going through 40m cavity was measured.

According to the fitting the coupling efficiency to the cavity is 98.46%, but still the beam looks loosely focused.

This measurement has been done by using the oplev legs (entry #2957) to allow the beam to go through the 40m walkway.

With a beam scan set on a movable cabinet, I measured it along the 40m chamber.

Since the plot looks not so nice,  I am going to work on this measurement a little bit more after I improve the mode matching.

 


Here is the parameters from the fitting

target waist [mm] 2.662
measured waist x [mm] 2.839
measured waist x [mm] 3.111
   
target waist position [m] 43.498
measured waist position x [m] 42.579
measured waist position y [m] 38.351

 

I believe the error for the travel length was within 0.5 meter. The length was always measured by a tape measure.

A thing I found was that: spatial jittering of the beam gets bigger as the beam goes further. This is the main source of the error bar for the spot size. 

 

 

Attachment 1: MMT40mcavity.png  69 kB  | Hide | Hide all
MMT40mcavity.png