Let me remind you how to lock and align the IMC
To lock
1. Open the doors for the IMC/OMC chambers. Open the manual shutter of the PSL just in front of the optical window
2. Run scripts/MC/mcloopson
3. Set the MC length path gain 0.3 / Set the MC total gain "+20"
4. If you want to avoid excitation of the mirrors by air turbulence, put a big plastic film and put three posters on the top and both the sides on the floor to block the wind go into the chamber.
To shut down
1. Run scripts/MC/mcloopsoff
2. Close the manual shutter, Remove the wind blockers, and the light door of the chambers
To align the MC
1. Tweak MC2 and MC3 to get maximum transmittion and/or minimum reflection. |