Entry  Mon May 10 16:59:35 2010, Alberto, Update, 40m Upgrading, Unexpected oscilaltionin the POY11 PD 2010-05-10_POY11_CalibratedOpticalResponse0-500MHz.pngPOX11.png
    Reply  Mon May 10 18:56:53 2010, rana, Update, Electronics, Unexpected oscilaltionin the POY11 PD 
       Reply  Mon May 10 19:09:45 2010, rana, Update, Electronics, Unexpected oscilaltionin the POY11 PD 
Message ID: 2905     Entry time: Mon May 10 19:09:45 2010     In reply to: 2904
Author: rana 
Type: Update 
Category: Electronics 
Subject: Unexpected oscilaltionin the POY11 PD 

Quote:

Where did you get the 55nH based notch from? I don't remember anything like that from the other LSC PD schematics. This is certainly a bad idea. You should remove it and put the notch back over by the other notch.

 Why is it a bad idea?

You mean putting both the 2-omega and the 55MHz notches next to each other right after the photodiode?