Message ID: 278
Entry time: Sun Jan 27 21:44:48 2008
|
Author: |
rana |
Type: |
Update |
Category: |
CDS |
Subject: |
Seismic BLRMS on Matlab |
|
|
I wrote a matlab script to produce band limited RMS trends from our accelerometers. It mimics the code written
by Ed Daw which makes the seismic FOMs at the sites.
Here's how it works:
- Use mDV to get data by reading directly from frames.
- Use the Matlab pwelch function to produce a power spectrum of the channels.
- Use the Matlab find function and rms.m to get the RMS in user-defined frequency bands.
- Makes a tdswrite command string which writes all the values to EPICS channels.
- The EPICS channels are just a list of simple names in a database file.
- The channel names are (will be) added to the C0EDCU.ini file so that its all trended.
The code is in the mDV/extra/C1 directory; its ~20 lines of code (excluding comments and spaces).
Next up is to add more DMF trend channels to the database and upgrade the code to use a .conf file
instead of hardcoded channel names. We should also evaluate if the bands I used are appropriate for the 40m;
I just used Ed's choices (0.1-0.3, 0.3-1, 1-3, 3-10, and 10-30 Hz).
In the medium term, we should make this compiled (like what RW did with the linetracker), and explore if we
want it to write values faster than 1/minute. |
|
% Seismic BLRMS Monitor
%
%
%
% RA 08-01-26
% 0 for no messages, 1 for debugging
debug_flag = 1;
% ------------ Build channel list