Entry  Thu Apr 1 00:43:33 2010, Mott, Update, General, PZT response for the innolight Innolight_Bode.png
    Reply  Thu Apr 1 07:17:15 2010, Koji, Update, General, PZT response for the innolight 
       Reply  Thu Apr 1 10:21:58 2010, Mott, Update, General, PZT response for the innolight Innolight_Response.pngLightwave_response.png
          Reply  Thu Apr 1 10:47:48 2010, Koji, Update, General, PZT response for the innolight 
             Reply  Thu Apr 1 12:07:22 2010, rana, Update, General, PZT response for the innolight 
                Reply  Thu Apr 1 18:05:29 2010, Mott, Update, General, PZT response for the innolight Innolight_Response_calFix.pngLightwave_response_calFix.png
                   Reply  Thu Apr 1 18:44:40 2010, Koji, Update, General, PZT response for the innolight 
                      Reply  Thu Apr 1 19:59:32 2010, Mott, Update, General, PZT response for the innolight Innolight_AM_Response.pngInnolight_AM_PM.pngInnoVsLW_PM.pngInnolight_AM_Response.pngLightwave_AM_PM.png
                         Reply  Mon Apr 12 14:20:10 2010, kiwamu, Update, Green Locking, PZT response for the innolight 
                            Reply  Tue Apr 13 19:53:06 2010, Mott, Update, Green Locking, PZT response for the innolight and lightwave 6x
Message ID: 2754     Entry time: Thu Apr 1 18:05:29 2010     In reply to: 2750     Reply to this: 2755
Author: Mott 
Type: Update 
Category: General 
Subject: PZT response for the innolight 

 

 We realized that we had measured the wrong calibration value; we were using the free-running error signal with the marconi far from the beat frequency, which was very small.  When we put the Marconi right at the beat, the signal increased by a factor of ~12 (turning our original calibration of 10 mV/rad into 120 mV/rad).  The re-calibrated plots are attached. 

Attachment 1: Innolight_Response_calFix.png  16 kB  | Show | Hide all | Show all
Attachment 2: Lightwave_response_calFix.png  18 kB  | Hide | Hide all | Show all
Lightwave_response_calFix.png
ELOG V3.1.3-