Thu Mar 18 10:46:51 2010, Koji, Update, ABSL, PLL reconstructed
|
Thu Mar 18 12:27:56 2010, Alberto, Update, ABSL, PLL reconstructed
|
Thu Mar 18 13:40:35 2010, Koji, Update, ABSL, PLL reconstructed
|
Thu Mar 18 21:42:26 2010, Koji, Update, ABSL, PLL reconstructed
|
Mon Mar 22 22:11:26 2010, Mott, Update, ABSL, PLL reconstructed
|
Mon Mar 22 23:37:32 2010, Mott, Update, ABSL, PLL reconstructed 
|
Tue Mar 23 18:44:46 2010, Mott, Update, ABSL, PLL reconstructed
|
Wed Mar 24 03:58:18 2010, kiwamu, matt, koji, Update, Green Locking, leave PLL locked
|
Wed Mar 24 04:22:51 2010, kiwamu, matt, Update, Green Locking, two NPRO PLL
|
Wed Mar 24 14:52:02 2010, Alberto, Update, Green Locking, two NPRO PLL
|
Wed Mar 24 15:59:59 2010, kiwamu, matt, Update, Green Locking, leave PLL locked
|
Thu Mar 25 17:29:48 2010, kiwamu, mott, Update, Green Locking, PLL two NPROs
|
|
Message ID: 2679
Entry time: Thu Mar 18 10:46:51 2010
Reply to this: 2680
|
Author: |
Koji |
Type: |
Update |
Category: |
ABSL |
Subject: |
PLL reconstructed |
|
|
Last night (Mar 17) I checked the PLL setup as Mott have had some difficulty to get a clean lock of the PLL setting.
- I firstly found that the NPRO beam is not going through the Faraday isolator well. This was fixed by aligning the steering mirrors before the Faraday.
- The signal from the RF PD was send to the RF spectrum analyzer through a power splitter. This is a waist of the signal. It was replaced to a directional coupler.
- Tee-ing the PZT feedback to the oscilloscope was producing the noise in the laser frequency. I put the oscilloscope to the 600Ohm output of the SR560, while connectiong the PZT output to the 50Ohm output.
- In addition, 6dB+6dB attenuators have been added to the PZT feedback signal.
Now the beating signal is much cleaner and behave straight forward. I will add some numbers such as the PD DC output, RF levels, SR560 settings...
Now I am feeling that we definitely need the development of really clean PLL system as we use PLL everywhere! (i.e. wideband PD, nice electronics, summing amplifiers, stop poking SR560, customize/specialize PDH box, ...etc) |