40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Sun Feb 28 23:47:52 2010, rana, Update, Computers, rosalba 
    Reply  Mon Mar 1 22:38:12 2010, rana, Update, Computers, RC sensitivity to RIN Untitled.png
Message ID: 2646     Entry time: Sun Feb 28 23:47:52 2010     Reply to this: 2649
Author: rana 
Type: Update 
Category: Computers 
Subject: rosalba 

Since Rosalba wanted to update ~500 packages, I let it do it. This, of course, stopped the X server from running. I downloaded and installed the newest Nvidia driver and its mostly OK.

The main problem with the auto-update on our workstations is that we've updated some packages by hand; i.e. not using the standard CentOS yum. So that means that the auto-update doesn't work right. From now on, if you want to install a fancier package than what CentOS distributes, you should commit to handle the system maintenance for these workstations for the future. Its not that we can't have new programs, we just have to pay the price.

 

At 23:45 PST, I also started a slow triangle wave on the AOM drive amplitude. This is to see if there's a response in the FSS-FAST which might imply a coupling from intensity noise to frequency noise via absorbed power and the dn/dT effect in the coatings.

Its a 93 second period triangle modulating the RC power from 100% down to 50%.