In this night, I checked the free swinging spectra of ETMX to make sure nothing wrong with ETMX by the wiping.
Compared with the past (Aug.6 2008), the spectra of ETMX doesn't show significant change.
Successfully the wiping activity didn't change its configuration so much and didn't bring bad situations.
(bad situation means for example, the suspended components hit some others).
The spectra of ETMX by DTT are attached. Also you can see the past spectra in Yoichi's entry.
Yoichi's data was taken during the air-pressure condition, so it's good for comparing.
Actually I compared those data by my eyes, because I could not get the past raw data somehow.
The resonant frequencies and their typical height changed a little bit, but I think those are not significant.
NOTE: In the figure, pitch and yaw modes (~0.57Hz and ~0.58Hz) look like having a smaller Q-factor than the past.
|