Message ID:
23
Entry time:
Mon Oct 29 09:16:31 2007
Author:
steve
Type:
Routine
Category:
VAC
Subject:
the rga is back
We had no filament current since last power glitch of Oct. 8, 2007
First I thought that the filament was lost, but it was only bad contact.
The rga head pins were oxidized. Rga was turned back on last Friday.
It's temp is 55.3C normal
Attachment 1:
rgaisback.jpg
273 kB
|
Hide
|