40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Nov 4 23:17:51 2009, Alberto, Update, LSC, X Arm Cavity transfer Function ArmCavityTF02.png
    Reply  Thu Nov 5 05:07:22 2009, rana, Update, LSC, X Arm Cavity transfer Function 
       Reply  Thu Nov 5 19:25:11 2009, Alberto, Update, LSC, X Arm Cavity transfer Function PDA255CalibrationTF03.png
Message ID: 2184     Entry time: Thu Nov 5 19:25:11 2009     In reply to: 2178
Author: Alberto 
Type: Update 
Category: LSC 
Subject: X Arm Cavity transfer Function 

Quote:

I would have guessed that you have to calibrate the detectors relative to each other before trying this. Its also going to be tricky if you use 2 different kinds of ADC for this (c.f. today's delay discussion in the group meeting).

I think Osamu used to look at fast transmission signals by making sure the PD at the end had a 50 Ohm output impedance and just drive the 40m long cable and terminate the receiving end with 50 Ohms. Then both PDs go into the SR785.

 

On Rana's suggestion I measured the trasfer function between the two photodiodes PDA255 that I'm using.

I took the one that I had on the end table and put it on the PSL table. I split the MC transmitted beam with a 50% beam splitter and sent the beams on the two diodes. (Rana's idea of picking off the beam and interposing the PDs before the ISS PDs was not doable: ISS PDs would be too close and there would be no room to install the PDA255 before them). See picture with the final setup.

DSC_0958.JPG


The transfer function also includes the 40m long cable that I was using for the Arm Cavity measurement.

Here's what I got. It looks rather flat. Yesterday the calibration was probably not the problem in that measurement.

PDA255CalibrationTF03.png
 I'm now going to install the PD back on the end table and measure the TFs between the excitation and several points of the loop.

(Trivia. At first, the PDs were saturating so Koji attached attenuation filters on to them. Suddenly the measurement got much nicer)

ELOG V3.1.3-