I remeasured the OAF time delay using the OAF-TF template from the Templates/ directory.

Troublingly, I found the MC1 dewhitening switches set OFF - please make sure that the MC1 dewhitening is back ON after each OAF tuning so that the interferometer locking is not hosed.

The OAF-TF template had the excitation amplitude set ~20x too high. I reduced it and the coherence was still > 0.95. The phase at 32 Hz was still ~126 deg as Jenne had measured, but since the phase at DC is 180 deg, the overall phase lag is just 180-126 = 54 deg. So the delay should be 54/180 * 32 = 9.7 => 10. Luckily, Jenne is working on an instructional manual for OAF that will make all of this crystal clear.