40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Sep 4 16:09:26 2009, josephb, Summary, VAC, Rebooted RGA computer and reset RGA settings 
    Reply  Tue Jan 21 17:26:59 2014, Jenne, Summary, VAC, Rebooted RGA computer and reset RGA settings 
       Reply  Wed Jan 22 15:24:11 2014, Steve, Summary, VAC, Rga scan after reboot Rgascan169dwarmingup.pngBlankMedmMustBeFixed.png
Message ID: 1967     Entry time: Fri Sep 4 16:09:26 2009     Reply to this: 9562
Author: josephb 
Type: Summary 
Category: VAC 
Subject: Rebooted RGA computer and reset RGA settings 

Steve noticed the RGA was not working today.  It was powered on but no other lights were lit.

Turns out the c0rga machine had not been rebooted when the file system on linux1 was moved to the raid array, and thus no longer had a valid mount to /cvs/cds/.  Thus, the scripts that were run as a cron could not be called.

We rebooted c0rga, and then ran ./RGAset.py to reset all the RGA settings, which had been reset when the RGA had lost power (and thus was the reason for only the power light being lit).

 

Everything seems to be working now.  I'll be adding c0rga to the list of computers to reboot in the wiki.