Entry  Thu Aug 6 23:29:47 2009, Jenne, Update, PSL, Ref cav reflection PD is funky 
    Reply  Fri Aug 7 00:10:14 2009, rana, Update, PSL, Ref cav reflection PD is funky 
       Reply  Fri Aug 7 17:05:34 2009, Jenne, Update, PSL, Ref cav reflection PD is funky 
Message ID: 1851     Entry time: Fri Aug 7 00:10:14 2009     In reply to: 1850     Reply to this: 1860
Author: rana 
Type: Update 
Category: PSL 
Subject: Ref cav reflection PD is funky 

we have a tester cable, but you don't want it. Instead the problem is probably at the cross-connect. The D-cable goes to a cross-connect and you can probe there with a voltmeter. If the signal is good there, trace it to the ADC. Also trend for several years to see when this happened - Yoichi may know the history better.

Also, we still need to complete the FSS RFPD task list from last year.

 