40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Thu Aug 6 17:51:21 2009, Chris, Update, General, Displacement Sensor Update Actual_Sensor.pngSensitivity.png
    Reply  Thu Aug 6 18:21:03 2009, Chris, Update, General, Displacement Sensor Update 
Message ID: 1845     Entry time: Thu Aug 6 17:51:21 2009     Reply to this: 1846
Author: Chris 
Type: Update 
Category: General 
Subject: Displacement Sensor Update 

For the past week Dmass and I have been ordering parts and getting ready to construct our own modified version of EUCLID (figure).  Changes to the EUCLID design could include the removal of the first lens, the replacement of the cat's eye retroreflector with a lens focusing the beam waist on a mirror in that arm of the Michelson, and the removal of the linear polarizers.  A beam dump was added above the first polarizing beam splitter and the beam at Photodetector 2 was attenuated with an additional polarizing beam splitter and beam dump.  Another proposed alteration is to change the non-polarizing beam splitter from 50/50 to 33/66.  By changing the reflectivity to 66\%, less power coming into the non-polarizing beam splitter would be ``lost" at the reference detector (1/3 instead of 1/2), and on the return trip less power would be lost at the polarizing beam splitter (1/6 instead of 1/4).  Also, here's a noise plot comparing a few displacement sensors that are used to the shot noise levels for the three designs I've been looking at.

Attachment 1: Actual_Sensor.png  107 kB  | Hide | Hide all | Show all
Actual_Sensor.png
Attachment 2: Sensitivity.png  132 kB  | Show | Hide all | Show all
ELOG V3.1.3-