Tue Jul 21 18:46:47 2009, steve, Configuration, Computers, computers are down
|
Wed Jul 22 01:57:19 2009, Alberto, Configuration, Computers, computers are down
|
Wed Jul 22 09:04:10 2009, Alberto, Configuration, Computers, computers are down
|
Wed Jul 22 11:14:11 2009, Alberto, Configuration, Computers, computers are down
|
Wed Jul 22 11:18:49 2009, rob, Configuration, Computers, sticky sliders
|
|
Message ID: 1776
Entry time: Wed Jul 22 11:14:11 2009
In reply to: 1773
Reply to this: 1777
|
Author: |
Alberto |
Type: |
Configuration |
Category: |
Computers |
Subject: |
computers are down |
|
|
Quote: |
Quote: |
Quote: |
All suspentions are kicked up. Sus dampings and oplev servos turned off.
c1iscey and c1lsc are down. c1asc and c1iovme are up-down
|
The computers and RFM network are up working again. A boot fest was necessary.Then I restored all the parameters with burtgooey.
The mode cleaner alignment is in a bad state. The autolocker can't get it locked. I don't know what caused it to move so far from the good state that it was till this afternoon. I went tuning the periscope but the cavity alignment is so bad that it's taking more time than expected. I'll continue working on that tomorrow morning.
|
I now suspect that after the reboot the MC mirrors didn't really go back to their original place even if the MC sliders were on the same positions as before.
|
Alberto, Rob,
we diagnosed the problem. It was related with sticky sliders. After a reboot of C1:IOO the actual output of the DAC does not correspond anymore to the values read on the sliders. In order to update the actual output it is necessary to do a change of the values of the sliders, i.e. jiggling a bit with them. |