Message ID: 1754     Entry time: Wed Jul 15 18:35:11 2009
Author: Stephanie 
Type: Update 
Category: General 
Subject: Multiply Resonant EOM Update 

Using FET probes, I was able to measure a transfer function that looks a little more like what I expected. There are only two peaks, but I think this can be explained by a short between the two capacitors (and two tunable capacitors) in the LC pairs, as shown (in red) in the circuit diagram attached. The measured transfer function (black), along with the simulated transfer functions with (red) and without (blue) the short are shown in the attached plot. The measured transfer function doesn't look exactly like the simulated transfer function with the short, but I think the difference can be explained by stray impedances.

Attachment 1: BuiltCkt1_Final.png  29 kB  | Hide | Hide all
BuiltCkt1_Final.png
Attachment 2: BuiltCkt1_TransferFunctions.png  12 kB  | Hide | Hide all
BuiltCkt1_TransferFunctions.png