Message ID: 1716     Entry time: Thu Jul 2 17:37:36 2009
Author: steve 
Type: Update 
Category: VAC 
Subject: V1 interlock test 

Joe, Alberto and Steve

We tested gate valve V1 interlock by :

1, decelerated rotation by brake from maglev controller unit.

2, turned maglev  controller off from controller unit.

3, unpluged 220VAC plug from wall socket

None of the above action triggered V1 to close. This needs to be corrected in the future.

The MEDM monitor screen of maglev indicated the correct condition changes.

 

 