Message ID: 17101     Entry time: Wed Aug 24 10:49:43 2022
Author: Cici 
Type: Update 
Category: General 
Subject: Measuring DFD output/X-arm laser PZT TF with Moku 

We measured the TF of the X-arm laser PZT using the Moku so we can begin fitting to that data and hopefully creating a digital filter to cancel out PZT resonances. 

-------------------------------------------------------------

We calculated the DFD calibration (V/Hz) using:

Vrf = 0.158 mV (-6 dBm), Km = 1 (K_phi = Km*Vrf), cable length = 45m,  Tau = cable length/(0.67*3*10^8 m/s) ~ 220 ns. 

We've taken some preliminary data and can see the resonances around 200-300 kHz.

---------------------------------------------------------

Next steps are taking more data around the resonances specifically, calibrating the data using the DFD calibration we calculated, and adjusting parameters in our model so we can model the TF.

 

Attachment 1: AUX_PZT_Actuator_nofit.pdf  35 kB  Uploaded Wed Aug 24 14:11:11 2022  | Hide | Hide all