[Koji, JC]
Koji and I have prepared the vacuum system for the power outage on Saturday.
- Closed V1 to isolate the main volume.
- Closed of VASE, VASV, VABSSCI,VABS, VABSSCO, VAEV, and VAEE.
- Closed V6, then close VM3 to isolate RGA
- Turn off TP1 (You must check the RPMs on the TP1 Turbo Controller Module)
- Close V5
- Turn off TP3 (There is no way to check the RPMs, so be patient)
- Close V4 (System State changes to 'All pneumatic valves are closed)
- Turn off TP2 (There is no way to check the RPMs, so be patient)
- Close Vacuum Valves (on TP2 and TP3) which connect to the AUX Pump.
- Turn of AUX Pump with the breaker switch wall plug.
From here, we shutdown electronics.
- Run
/sbin/shutdown -h now on c1vac to shut the host down.
- Manually turn off power to electronic modules on the rack.
- GP316a
- GP316b
- Vacuum Acromags
- PTP3
- PTP2
- TP1
- TP2 (Unplugged)
- TP3 (Unplugged)
|